Why these products were recalled

Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (BRF) modules with software version 02-xx, only when rack buffering is activated and the rack buffer is completely full. This issue can lead to sample spillage, posing a potential risk to operators / laboratory staff due to exposure to potentially infectious material, cross-contamination of samples in the affected racks and erroneous results due to sample carry-over.

"Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (URF) modules with software version 02-xx, only when rack buffering is activated and the rack buffer is completely full. This issue can lead to sample spillage, posing a potential risk to operators / laboratory staff due to exposure to potentially infectious material, cross-contamination of samples in the affected racks and erroneous results due to sample carry-over."

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2017-04-19

    Roche Recalls Cobas 8100 Reformatter Modules Due to Rack Crash Risk

    Roche Diagnostics is voluntarily recalling 45 Cobas 8100 reformatter modules because a rack crash can cause sample spillage, posing a risk of exposure to infectious material.

    Product
    Cobas 8100 uni-directional reformatter (BRF) module with Software Version 02-xx
    Category
    Distribution
    16 states
  • HighFDA (Devices)··2017-04-19

    Roche Recalls Cobas 8100 Modules Due to Potential Sample Spillage

    Roche Diagnostics is voluntarily recalling 45 Cobas 8100 reformatter modules because a software issue can cause rack crashes and sample spillage, posing a risk of exposure to infectious material.

    Product
    Cobas 8100 bi-directional reformatter (BRF) module with Software Version 02-xx
    Category
    Distribution
    16 states