The Recall Desk

FDA (Devices) recall action 70322

Navilyst Medical, Inc recalled 60 products on 2015-03-04

The FDA (Devices) publishes a recall like this one as 60 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 60 pages.

Products
60
Announced
2015-03-04
Critical
0
Units
16,508

Why these products were recalled

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–50 of 60

  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 5 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, LEFT HEART KIT, UPN H749601506721, REF/Catalog No. 60150672, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitorin
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 2 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, LEFT HEART KIT, UPN H749601605351, REF/Catalog No. 60160535, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 10 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, ANGIO KIT, UPN H749601319961, REF/Catalog No. 60131996, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136683. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, 3V MANIFOLD KIT, UPN H965601011621, REF/Catalog No. 60101162, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 260 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, UPN H965600338021, REF/Catalog No. 60033802, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 2 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, LEFT HEART KIT, UPN H965600125911, REF/Catalog No. 60012591, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, UPN H749601399921, REF/Catalog No. 60139992, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 5 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, LEFT HEART KIT, UPN H965600416321, REF/Catalog No. 60041632, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 3 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, LEFT HEART, POTENTIAL REVISION TO 600604412, UPN H965PK0604436S1, REF/Catalog No. PK0604436S, STERILE, Rx ONLY - For single use only. --- Device Listing Number D136588 NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pr
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 64 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749601014911, REF/Catalog No. 60101491, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136683. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitori
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 11 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, HOPKINS LEFT HEART, UPN H749601836621, REF/Catalog No. 60183662, STERILE, Rx ONLY ---- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 136 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, UPN H749600329421, REF/Catalog No. 60032942, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 17 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, LEFT HEART KIT, UPN H749602207461, REF/Catalog No. 60220746, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 512 NAMIC Convenience Kits because expiration dating is not fully supported by testing, posing a potential sterility breach and infection risk.

    Product
    NAMIC Convenience Kit, DYE CATH LAB KIT, UPN H749601367131, REF/Catalog No. 60136713, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monito
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 69 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, LEFT HEART KIT, UPN H749601617931, REF/Catalog No. 60161793, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 438 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, 3 VALVE KIT CARDIOVASCULAR PROCEDURE KIT, UPN H749600403131, REF/Catalog No. 60040313, STERILE, Rx ONLY --- --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst NAMIC Convenience Kit Recalled Due to Sterility Concerns

    Navilyst Medical is recalling 602 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, 3 VALVE KIT, UPN H749600709971, REF/Catalog No. 60070997, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, UPN H7496016046121, REF/Catalog No. 601604612, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 74 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, UPN H965601931651, REF/Catalog No. 60193165, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 30 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, RADIOLOGY KIT, UPN H749600617521, REF/Catalog No. 60061752, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136683. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 2,502 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, UPN H749602301481, REF/Catalog No. 60230148, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 124 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, UPN H7496013148131, REF/Catalog No. 601314813, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 19 units of NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, 5 VALVE MANIFOLD KIT, UPN H749600105121, REF/Catalog No. 60010512, STERILE, Rx ONLY --- For single use only. Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monito
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 16 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased risk of infection.

    Product
    NAMIC Convenience Kit, MOBILE CATH LAB KIT, UPN H749600419211, REF/Catalog No. 60041921, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Mon
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2015-03-04

    Navilyst Medical Recalls NAMIC Convenience Kits Due to Sterility Concerns

    Navilyst Medical is recalling 541 NAMIC Convenience Kits because expiration dating is not fully supported by testing, which may result in a sterility breach and increased infection risk.

    Product
    NAMIC Convenience Kit, UPN H965607011741, REF/Catalog No. 60701174, STERILE, Rx ONLY --- For single use only. --- Device Listing Number D136588. NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
    Category
    Distribution
    0 states