The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

12626–12650 of 13837

  • HighFDA (Devices)·Z-0289-2022·2021-12-01

    CT Scanner Software May Cause Workflow Delays and Patient Rescans

    Siemens SOMATOM Edge Plus CT scanners with syngo.CT VB20 software may experience workflow interruptions, potentially delaying diagnoses and requiring patient rescans. The issue affects 64 units nationwide.

    Product
    SOMATOM Edge Plus with software syngo.CT VB20 Model #10267000
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0293-2022·2021-12-01

    Medtronic Wireless Recharger for InterStim Micro May Become Unresponsive

    Medtronic's wireless recharger for the InterStim Micro implantable neurostimulator may become unresponsive if charging procedures from the user guide are not followed. Approximately 10,081 devices are affected nationwide.

    Product
    Medtronic Wireless Recharger Model WR9220 included in the RS5200 Recharger Kit used by the InterStim Micro (Model 97810) implantable neurostimulator.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0295-2022·2021-12-01

    Medtronic Wireless Recharger Model WR9200 Becomes Unresponsive on Improper Use

    Medtronic is recalling the Wireless Recharger Model WR9200 used with RestoreUltra and RestoreSensor neurostimulator implants. The recharger becomes unresponsive if charging instructions from the user guide are not followed correctly.

    Product
    Medtronic Wireless Recharger Model WR9200 included in the RS7200 Recharger Kit used by the RestoreUltra (Model 37712), RestoreSensor (Model 37714), Restore Ultra SureScan MRI (Model 97712), and RestoreSensor SureScan MRI (Model 97714) implantable neurostimulators.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0292-2022·2021-12-01

    Medtronic Endurant IIs stent graft recalled due to spindle detachment risk

    Medtronic recalls 419 Endurant IIs stent grafts due to risk of spindle detachment from the hypotube during deployment.

    Product
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF281
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0282-2022·2021-12-01

    Medtronic Cardiac Device Software Affected by Data Processing Error

    Medtronic is recalling software used in certain cardiac implant devices due to a data processing error that may fail to collect episode data. Approximately 18,715 devices are affected worldwide.

    Product
    Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds): Cobalt XT VR: DVPA2D1, DVPA2D4; Cobalt VR:
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0284-2022·2021-12-01

    CT Scanner Software Issue Causes Workflow Interruptions and Diagnostic Delays

    A software issue in SOMATOM Definition AS CT scanners may cause workflow interruptions, leading to diagnostic delays and the need for patient rescans. The recall affects approximately 665 units distributed nationwide.

    Product
    SOMATOM Definition AS with software syngo.CT VB20 Model #8098027
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0240-2022·2021-12-01

    Medical X-Ray Systems Missing Required Radiation Warning Labels

    Philips Healthcare is recalling ProxiDiagnost N90 diagnostic X-ray systems that lack required FDA radiation warning labels. Some units nationwide were found missing the mandatory radiation safety labels and certification statements required by federal standards.

    Product
    ProxiDiagnost N90 is multi-functional general R/F systems.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0304-2022·2021-12-01

    Siemens Artis Q.zen Angiography System X-ray Activation Failure

    Siemens is recalling 15 Artis Q.zen angiography systems that may fail to generate x-rays when the lift mechanism and x-ray release are activated simultaneously, potentially causing a procedural delay.

    Product
    Artis Q.zen ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848354
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0302-2022·2021-12-01

    Siemens Artis Q Biplane X-ray System Software Imaging Delay Issue

    Siemens Medical Solutions is recalling 15 Artis Q biplane diagnostic x-ray systems due to a software issue in the VD12-angiography subsystem. When source-to-image distance lift is activated simultaneously with x-ray release, the system fails to activate and displays an error message, causing procedural delay.

    Product
    Artis Q biplane with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848282
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0275-2022·2021-12-01

    da Vinci Surgical Drapes Recalled for Incorrect Expiration Date on Packaging

    Intuitive Surgical recalls 1,740 da Vinci Instrument Arm Drapes due to incorrect expiration dates on package labeling. The error may prevent accurate verification of sterility assurance dates.

    Product
    The da Vinci Xi and X surgical systems (IS4000 and IS4200) each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the system and its primary function is to support the instru
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0297-2022·2021-12-01

    Siemens Artis zee x-ray system may delay imaging delivery during procedures

    Siemens is recalling Artis zee ceiling-mounted x-ray imaging systems equipped with VD12 software. A software issue may prevent x-ray delivery when the source-to-image distance lift is activated during procedures, resulting in procedural delays.

    Product
    Artis zee ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number:10094137
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0296-2022·2021-12-01

    Medical Imaging System May Fail to Release X-rays Under Specific Conditions

    Siemens Artis zee floor angiography systems may fail to release x-rays when the SID lift and x-ray functions are activated simultaneously, causing procedural delay. Three units are affected.

    Product
    Artis zee floor with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10094135
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0298-2022·2021-12-01

    Siemens Artis Zee Angiography System Software May Prevent X-Ray Activation

    Siemens is recalling 19 units of its Artis zee angiography system due to a software issue that may prevent x-ray activation under specific operating conditions, causing procedural delay.

    Product
    Artis zee multi-purpose with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10094139
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0303-2022·2021-12-01

    Siemens Artis Q.zen Angiography System May Fail to Fire X-rays During Simultaneous Operations

    Three Siemens Artis Q.zen angiography systems may fail to fire X-rays when the source-to-image distance lift is activated simultaneously with X-ray commands. This causes a procedural delay.

    Product
    Artis Q.zen floor with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848353
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0299-2022·2021-12-01

    Siemens Artis Zee Biplane X-ray Imaging System Software Malfunction

    Siemens Artis zee biplane X-ray imaging systems may fail to release X-rays when the source-to-image distance lift is activated simultaneously, resulting in procedural delays.

    Product
    Artis zee biplane with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10094141
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0305-2022·2021-12-01

    Siemens Artis Q.zen Angiography System X-Ray Imaging Delay Issue

    Siemens has recalled two Artis Q.zen angiography systems due to a software issue that may prevent x-ray imaging and delay procedures when SID lift and x-ray controls are activated simultaneously.

    Product
    Artis Q.zen biplane with software VD12- angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number:10848355
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0300-2022·2021-12-01

    Siemens Artis Q floor angiography system: X-ray failure during SID lift operation

    Siemens Artis Q floor angiography systems may fail to produce x-ray when SID lift movement is activated during x-ray release, resulting in procedural delays. The system displays an error message.

    Product
    Artis Q floor with software VD12- angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848280
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0233-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device inner lumen detachment and embolization risk

    Arrow-Trerotola PTD catheters are being recalled because the inner lumen can detach from the device basket, causing embolization that requires additional intervention. 1,167 units are affected.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-W The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venousf
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0234-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device: Inner Lumen Detachment Risk

    Arrow International is recalling 1918 units of the Arrow-Trerotola Percutaneous Thrombolytic Device due to inner lumen detachment that can cause embolization, requiring emergency retrieval and delaying treatment.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-WC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0232-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device Catheter Lumen Detachment Recall

    Arrow International is recalling Arrow-Trerotola Percutaneous Thrombolytic Device catheters because the inner lumen can detach from the device basket, causing embolization requiring additional intervention and delaying dialysis therapy. Approximately 1,245 units are affected worldwide.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-HFWC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-veno
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0231-2022·2021-11-24

    Arrow-Trerotola Percutaneous Thrombolytic Device Inner Lumen Detachment Recall

    The inner lumen of Arrow-Trerotola Over-The-Wire PTD catheters (7FR) can detach from the device, causing an embolism. This requires additional medical intervention and delays treatment.

    Product
    Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-12709-WC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venous
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0224-2022·2021-11-24

    FDA Recalls Scrip Ultrasound Gel for Potential Bacterial Contamination

    Scrip Inc is recalling Scrip Ultrasound Gel Clear (5L) nationwide due to potential bacterial contamination. Affected units may cause harm if contaminated product contacts patients during ultrasound procedures.

    Product
    Scrip Ultrasound Gel Clear, 5L Intended for use in acoustic coupling of ultrasound transducer to tissue
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-0225-2022·2021-11-24

    Scrip Creme ultrasound lotion recalled for bacterial contamination

    Scrip Creme All Purpose Lotion is being recalled due to bacterial contamination in this ultrasound coupling lotion. The product—1001 gallons in lots distributed between August 2018 and August 2021—was distributed nationwide.

    Product
    Scrip Creme All Purpose Lotion Intended for use in acoustic coupling of ultrasound transducer to tissue
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0255-2022·2021-11-24

    REGARD Ultrasound Trays Recalled for Out-of-Specification Swabstick Component

    ROi CPS LLC has recalled REGARD Ultrasound Trays because they contain out-of-specification Povidone-Iodine swabstick components. The 240 affected trays were distributed in Florida, Louisiana, and Missouri.

    Product
    REGARD Ultrasound Tray
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0266-2022·2021-11-24

    Cobas u 601 urinalysis analyzer recalled for false negative nitrite results

    Roche Diagnostics is recalling the Cobas u 601 urinalysis test system due to potential false negative nitrite results when endogenous creatinine levels exceed 15,000 mg/L. The device may fail to detect nitrite in urine samples with elevated creatinine.

    Product
    Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is comprised of the cobas u 601 urine analyzer and the cobas u pack. The cobas u 601 urine analyzer when used with the cobas u pack is a fully automated urinalysis system intend
    Category
    Medical Device
    Distribution
    Distributed nationwide