The Recall Desk

Manufacturer

Philips Medical Systems Nederland B.V.

209 recalls in our database name Philips Medical Systems Nederland B.V. as the manufacturing or recalling firm.

What the numbers show

Total
209
Critical
2
Severe
20
Most recent
2026-08-19

Of the 209 recalls from Philips Medical Systems Nederland B.V. we have scored against our rubric, 2 (1%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 20 (10%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

151–175 of 209

  • HighFDA (Devices)·Z-1077-2025·2025-02-12

    Philips AD7 and AD7X patient tables recalled due to finger entrapment risk

    Philips has recalled the AD7 and AD7X patient tables used in medical imaging systems because fingers can become entrapped between the tabletop and guiding rails during manual repositioning, potentially causing injury.

    Product
    Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1093-2025·2025-02-12

    Philips Azurion patient tables recalled for finger entrapment risk

    Philips is recalling Azurion 7 M12 and Allura patient tables used in medical imaging systems due to a finger entrapment hazard during manual repositioning. Operators and service personnel may suffer finger injury if fingers are caught between the tabletop and guiding rails.

    Product
    Azurion 7 M12 System Code: (1) 722078 (2) 722223 (3) 722233
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1084-2025·2025-02-12

    Philips Allura and Azurion Patient Tables Recall Due to Finger Entrapment Risk

    Philips is recalling AD7 and AD7X patient tables used in Allura and Azurion medical imaging systems due to a finger entrapment hazard during manual repositioning. Operators and service personnel are at risk of finger injury.

    Product
    AlluraXperFD20/10 System Code: (1) 722029
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1086-2025·2025-02-12

    Philips AD7 and AD7X Patient Table Finger Entrapment Risk

    The Philips AD7 and AD7X patient tables pose a finger entrapment risk during manual repositioning. Operators and service personnel could sustain finger injuries from the gap between the longitudinal guiding rails and tabletop.

    Product
    AlluraXperFD20/15 System Code: (1) 722058
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1092-2025·2025-02-12

    Philips Azurion patient tables recalled for finger entrapment hazard

    Philips is recalling Azurion 7 B20 patient tables due to a finger entrapment hazard between the tabletop and guiding rails during manual repositioning. Operators and service personnel could suffer finger injury.

    Product
    Azurion 7 B20 System Code: (1) 722068 (2) 722226 (3) 722236
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1088-2025·2025-02-12

    Philips Azurion Patient Tables Recalled for Finger Entrapment Risk

    Philips Azurion patient tables used in medical imaging systems can trap fingers between the tabletop and rails during manual repositioning, potentially causing finger injury to operators and service personnel.

    Product
    Azurion 3 M15 System Code: (1) 722064 (2) 722222 (3) 722280
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1091-2025·2025-02-12

    Philips Azurion and Allura System Patient Tables: Finger Entrapment Risk

    Philips Azurion and Allura system patient tables may trap fingers between the guiding rails and tabletop during manual repositioning, risking finger injury to operators and service personnel.

    Product
    Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1081-2025·2025-02-12

    Philips AlluraXper FD20 Biplane Patient Table Finger Entrapment Risk

    Philips AlluraXper FD20 Biplane patient tables may cause finger entrapment between the tabletop and guiding rails during manual repositioning, affecting operators and service personnel.

    Product
    AlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1080-2025·2025-02-12

    Philips recalls AlluraXper FD20 patient tables due to finger entrapment hazard

    Philips is recalling AlluraXper FD20 patient tables due to a finger entrapment hazard. During manual repositioning, fingers can get trapped between the tabletop and rails, risking injury to operators and service personnel.

    Product
    AlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1082-2025·2025-02-12

    Philips AlluraXper FD20 operating room surgical table finger entrapment risk

    Philips AlluraXper FD20 operating room tables can trap fingers during manual repositioning, risking injury. The hazard affects both operators and service personnel worldwide.

    Product
    AlluraXper FD20 Biplane OR Table System Code: (1) 722020 (2) 722025
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1083-2025·2025-02-12

    Philips AlluraXper Operating Room Table Finger Entrapment Hazard

    The Philips AlluraXper FD20 operating room table can trap fingers between the guiding rails and tabletop during manual repositioning, potentially injuring operators and service personnel.

    Product
    AlluraXper FD20 OR Table System Code: (1) 722023 (2) 722035
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1076-2025·2025-02-12

    Philips Allura CV20 patient tables risk finger entrapment during repositioning

    The Philips Allura CV20 System patient tables (AD7 and AD7X models) pose a finger entrapment hazard during manual repositioning. Fingers can become caught between the tabletop and guiding rails, potentially injuring operators and service personnel.

    Product
    Allura CV20 System Code: 722031
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1087-2025·2025-02-12

    Philips Azurion and Allura Patient Tables: Finger Entrapment Hazard

    Philips patient tables on Azurion and Allura imaging systems can trap fingers between the tabletop and rails during manual repositioning. Operators and service personnel are at risk of finger injury.

    Product
    Azurion 3 M12 System Code: (1) 722063 (2) 722221
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1079-2025·2025-02-12

    Philips Allura patient tables recall due to finger entrapment risk

    Philips recalls AD7 and AD7X patient tables in Allura and Azurion systems due to finger entrapment risk during manual repositioning. Operators and service personnel could suffer finger injuries.

    Product
    Allura Xper FD10C System Code: (1) 722001
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1090-2025·2025-02-12

    Philips Azurion 5 M20 Patient Tables Recalled for Finger Entrapment Hazard

    Philips Azurion 5 M20 x-ray system patient tables can trap fingers between the tabletop and guiding rails during repositioning. The hazard may cause finger injury to operators and service personnel.

    Product
    Azurion 5 M20 System Code: (1) 722228 (2) 722232 (3) 722281
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1078-2025·2025-02-12

    Philips AlluraXper Patient Tables Pose Finger Entrapment Risk

    Philips is recalling AD7 and AD7X patient tables used in Allura and Azurion fluoroscopy systems. Manual repositioning can trap fingers between the tabletop and rails, risking injury to operators and service personnel.

    Product
    AlluraXper FD10/10 System Code: (1) 722005 (2) 722011 (3) 722027
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1094-2025·2025-02-12

    Philips Azurion and Allura Patient Table Finger Entrapment Risk

    Philips Azurion 7 M20 and Allura patient tables pose a finger entrapment risk during manual repositioning. Fingers can get caught between the tabletop and guiding rails, potentially causing finger injury to operators and service personnel.

    Product
    Azurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223 (3) 722233
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0676-2025·2024-12-18

    Philips Allura Xper X-ray Systems Prone to Clamping Bolt Failure and Collision Risk

    Philips Allura Xper radiological systems may experience propeller motor and x-ray tube locking bolt breaks, causing erratic C-arm movements with potential for collision with patients or bystanders, and unnecessary radiation exposure.

    Product
    Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0353-2025·2024-11-20

    Philips Zenition 70 Recall Due to Potential Fuse Failure

    Philips is recalling Zenition 70 devices because the mains control unit board fuse may blow during start-up or procedures.

    Product
    Philips Zenition 70, Model Number: 718133
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3165-2024·2024-09-25

    Philips Azurion X-ray Systems: Cable Hose Carrier Detachment Risk

    Philips recalls Azurion X-ray imaging systems due to potential cable hose carrier detachment from broken plastic or bolts, which may cause parts to fall and create injury risk.

    Product
    System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3167-2024·2024-09-25

    PHILIPS INTEGRIS X-ray systems cable hose carrier detachment recall

    PHILIPS medical X-ray imaging systems may have a cable hose carrier that can detach due to broken plastic or bolts, potentially causing parts to fall and injure clinical staff.

    Product
    System Name: 722044 INTEGRIS Allura 15-12 (biplane) 722030 Integris CV Cesar-Powerpack-Visub-Nicol 722021 INTEGRIS Allura 9 (biplane) 722018 Integris Allura 9 722017 Integris H5000 F / Allura 9F 722016 Integris H5000 C / Allura 9C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3168-2024·2024-09-25

    Philips MultiDiagnost Eleva X-ray system cable hose carrier detachment

    Philips is recalling 109 MultiDiagnost Eleva X-ray systems worldwide due to potential cable hose carrier detachment from broken plastic or bolts, which could cause parts to fall and result in injury.

    Product
    System Name: 708038 MultiDiagnost Eleva with Flat Detector 708037 MultiDiagnost Eleva with Flat Detector 708036 MultiDiagnost Eleva 708034 MultiDiagnost Eleva with Flat Detector 708032 MultiDiagnost Eleva
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3166-2024·2024-09-25

    Philips Allura Xper Fluoroscopy Systems Cable Hose Carrier Detachment

    Philips Allura Xper FD series fluoroscopy systems may have detached cable hose carriers due to broken plastic or bolts. A detachment could cause parts to fall and potentially result in injury.

    Product
    System Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 OR Table 722038 Allura Xper FD20/20 722035 Allura Xper FD20 OR Table 722029 Allura Xper FD20/10 722028 Allura Xper FD20 722027 Allura Xper FD10/10 722026 Allura Xper FD10…
    Category
    Medical Device
    Distribution
    Distributed nationwide