PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

192 recalls in our database name PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. as the manufacturing or recalling firm.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

151–175 of 192

  • HighFDA (Devices)·Z-1087-2025·2025-02-12

    Philips Azurion and Allura Patient Tables: Finger Entrapment Hazard

    Philips patient tables on Azurion and Allura imaging systems can trap fingers between the tabletop and rails during manual repositioning. Operators and service personnel are at risk of finger injury.

    Product
    Azurion 3 M12 System Code: (1) 722063 (2) 722221
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1088-2025·2025-02-12

    Philips Azurion Patient Tables Recalled for Finger Entrapment Risk

    Philips Azurion patient tables used in medical imaging systems can trap fingers between the tabletop and rails during manual repositioning, potentially causing finger injury to operators and service personnel.

    Product
    Azurion 3 M15 System Code: (1) 722064 (2) 722222 (3) 722280
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0676-2025·2024-12-18

    Philips Allura Xper X-ray Systems Prone to Clamping Bolt Failure and Collision Risk

    Philips Allura Xper radiological systems may experience propeller motor and x-ray tube locking bolt breaks, causing erratic C-arm movements with potential for collision with patients or bystanders, and unnecessary radiation exposure.

    Product
    Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3165-2024·2024-09-25

    Philips Azurion X-ray Systems: Cable Hose Carrier Detachment Risk

    Philips recalls Azurion X-ray imaging systems due to potential cable hose carrier detachment from broken plastic or bolts, which may cause parts to fall and create injury risk.

    Product
    System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3166-2024·2024-09-25

    Philips Allura Xper Fluoroscopy Systems Cable Hose Carrier Detachment

    Philips Allura Xper FD series fluoroscopy systems may have detached cable hose carriers due to broken plastic or bolts. A detachment could cause parts to fall and potentially result in injury.

    Product
    System Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 OR Table 722038 Allura Xper FD20/20 722035 Allura Xper FD20 OR Table 722029 Allura Xper FD20/10 722028 Allura Xper FD20 722027 Allura Xper FD10/10 722026 Allura Xper FD10 722025
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3168-2024·2024-09-25

    Philips MultiDiagnost Eleva X-ray system cable hose carrier detachment

    Philips is recalling 109 MultiDiagnost Eleva X-ray systems worldwide due to potential cable hose carrier detachment from broken plastic or bolts, which could cause parts to fall and result in injury.

    Product
    System Name: 708038 MultiDiagnost Eleva with Flat Detector 708037 MultiDiagnost Eleva with Flat Detector 708036 MultiDiagnost Eleva 708034 MultiDiagnost Eleva with Flat Detector 708032 MultiDiagnost Eleva
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3167-2024·2024-09-25

    PHILIPS INTEGRIS X-ray systems cable hose carrier detachment recall

    PHILIPS medical X-ray imaging systems may have a cable hose carrier that can detach due to broken plastic or bolts, potentially causing parts to fall and injure clinical staff.

    Product
    System Name: 722044 INTEGRIS Allura 15-12 (biplane) 722030 Integris CV Cesar-Powerpack-Visub-Nicol 722021 INTEGRIS Allura 9 (biplane) 722018 Integris Allura 9 722017 Integris H5000 F / Allura 9F 722016 Integris H5000 C / Allura 9C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2953-2024·2024-09-11

    Philips Azurion 7M20 Fluoroscopic X-Ray System FlexArm Movement Recall

    Philips Azurion 7M20 interventional fluoroscopic X-ray systems with FlexArm ceiling-mounted stands may experience inconsistent or unavailable longitudinal movement due to bearing grease leakage affecting device mobility. Approximately 927 systems worldwide are affected.

    Product
    Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2716-2024·2024-08-28

    Philips Allura Xper Interventional Fluoroscopy Systems Recalled for Incorrect Bolt

    Philips Allura Xper systems contain an incorrect half-threaded bolt in LTE kits instead of the required full-threaded bolt. Six units distributed worldwide have been affected.

    Product
    Philips Allura Xper systems (R8.1 with Poly-G Stand) Interventional Fluoroscopic X-Ray System, Model Numbers 722010, 722012, 722013
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2534-2024·2024-08-14

    Philips Azurion X-ray System: Potential Loss of Imaging Functionality

    The Philips Azurion Interventional Fluoroscopic X-ray System may lose imaging functionality or experience longer restart times. The FDA is recalling approximately 3,011 units of this device worldwide.

    Product
    Philips Azurion Interventional Fluoroscopic X-ray System with Software Version Number: R1.0, R1.1, R1.2- Intended for Image guidance in diagnostic, interventional, and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, car
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1897-2024·2024-06-05

    Philips Allura Xper FD20 Software Update Disables Pivot Brake on Patient Table

    A software upgrade to Philips Allura Xper FD20 Biplane systems causes the patient table pivot brake to fail to engage, potentially causing unexpected table movements during procedures.

    Product
    Philips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013, Software Version Number 8.1.100.
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1895-2024·2024-06-05

    Philips Allura Xper FD20 patient table brake may fail to engage after software update

    Philips Allura Xper FD20 X-ray systems with ADN7NT patient tables may have a pivot brake failure after software upgrade 8.1.100 is installed. This could cause unexpected table movements during use.

    Product
    Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1926-2024·2024-06-05

    Philips Hemodynamic Application software pressure wave data synchronization defect

    Philips Interventional Hemodynamic Application software versions 1.2.0–1.3.1 may not correctly synchronize pressure wave data when acquiring measurements from two devices simultaneously.

    Product
    Philips Interventional Hemodynamic Application R.1.2X, R.1.3.0, R1.2.1. Software that enables invasive investigation of cardiac and vascular diseases.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1896-2024·2024-06-05

    Philips Allura Xper FD10 system pivot brake failure after software update

    Philips is recalling Allura Xper FD10 medical imaging systems with software version 8.1.100 because the pivot brake may fail to engage, potentially causing unexpected table movements during procedures.

    Product
    Philips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010, Software Version Number 8.1.100.
    Category
    Medical Device
    Distribution
    1 state
  • HighFDA (Devices)·Z-1152-2024·2024-02-28

    Philips Azurion X-ray Systems Disk Bay May Fail and Prevent Imaging

    The disk bay component in certain Philips Azurion x-ray systems may fail, potentially preventing diagnostic imaging and delaying procedures. Affected units have been distributed worldwide.

    Product
    Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: XRay PC, Suite PC, and FlexViewing PC Azurion Model Name - Model Number: Azurion 3 M12 722063 722221; Azurion 3 M15 722064
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1176-2024·2024-02-28

    Philips Azurion 3 Surgical Imaging System Recalled for Power Failure Risk

    Philips is recalling 14 units of its Azurion 3 surgical imaging system due to a potential short circuit in the power inverter. The defect could cause the device to lose power and fail during surgical procedures.

    Product
    Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 722222 (3) 722280
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1177-2024·2024-02-28

    Azurion 5 X-Ray System Generator May Fail Due to Circuit Short

    Philips Azurion 5 X-ray systems with Certeray generators may experience power loss due to a potential short circuit in the power inverter board, potentially delaying or interrupting medical procedures.

    Product
    Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722227 (2) 722228
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1145-2024·2024-02-28

    Philips Azurion Interventional X-ray Systems framegrabber card may fail to display images

    The framegrabber card in Philips Azurion interventional X-ray systems may fail to display diagnostic images properly, potentially delaying procedures.

    Product
    Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected system. Affected PCs: FlexViewing PC Azurion Model Name - Model Number: Azurion 7 B12 722067; Azurion 7 B20 722068; Azurion 7 M12 722078 722223;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1178-2024·2024-02-28

    Azurion 7 Medical Imaging System May Lose Power During Procedures

    Philips Azurion 7 imaging systems may experience power loss due to a potential short circuit in the power inverter, risking delays or termination of diagnostic and surgical procedures.

    Product
    Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722223, (2) 722224, (3) 722225, (4) 722226
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1144-2024·2024-02-28

    Philips Allura Xper Interventional X-ray System Display and Imaging Malfunction

    A framegrabber card malfunction in Philips Allura Xper interventional X-ray systems may cause display failures, preventing proper visualization of medical images and delaying procedures.

    Product
    Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected systems. Affected PCs: FlexVision PC Allura Xper Model Name - Model Number: Allura Xper FD10 722003 722010 722026; Allura Xper FD10/10 722005
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1156-2024·2024-02-28

    Philips Allura Xper X-ray System Memory Modules May Stop Functioning

    Philips Allura Xper fluoroscopic X-ray systems may experience memory module failures that cause the system to stop functioning and prevent imaging. This malfunction could delay critical medical procedures.

    Product
    Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the DIMMs (Dual In-line Memory Modules) used with affected systems. Allura Xper Model Name - Model Number: Allura Xper FD10 722003 722010 722026; Allura Xper FD10/10 722005 722011 722027;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1151-2024·2024-02-28

    Philips Allura Xper X-ray systems disk bay component may malfunction

    Philips has recalled certain Allura Xper interventional X-ray systems because the disk bay component may fail, potentially preventing system operation and imaging procedures.

    Product
    Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected systems. Affected PCs: Imaging Processing PC, Host PC, FlexVision PC Allura Xper Model Name - Model Number: Allura Xper FD10 722003 722010 722026;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0860-2024·2024-02-07

    Philips Azurion 7 M20 FlexArm system recalled for software connectivity loss

    Philips Medical Systems recalls the Azurion 7 M20 FlexArm system due to a software issue causing loss of connectivity between the FlexArm and Maquet Magnus Table. The recall affects 28 systems worldwide.

    Product
    Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0835-2024·2024-02-07

    Philips Azurion X-ray System Software Issue Causes Restart and Imaging Loss

    Philips Azurion interventional x-ray systems may continuously restart due to a software issue, causing loss of imaging functionality and data. If this occurs during a procedure, the procedure may be delayed or aborted.

    Product
    Philips Azurion system, Interventional fluoroscopic x-ray system. System Product Name - Model Numbers Azurion 3M12 -722063, 722221; Azurion 3M15 - 722064, 722222; Azurion 5M12 - 722227; Azurion 5M20 - 722228; Azurion 7B12/12 - 722067, 722225; Azurion 7B20/15 - 722068, 722226;
    Category
    Medical Device
    Distribution
    Distributed nationwide