The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

6151–6175 of 40520

  • HighFDA (Devices)·Z-0113-2025·2024-10-23

    Boston Scientific AVVIGO+ Guidance System Grid Mark Display Error

    Boston Scientific is recalling the AVVIGO+ Guidance System due to a software bug displaying incorrect grid marks during live imaging. The device shows 9 grid marks instead of 15, which could affect accurate vessel measurement in cardiac interventions.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0122-2025·2024-10-23

    Medline Open Heart Surgical Kit Recall for Model Labeling Error

    Medline Industries recalls certain open-heart surgical kits (Lot 23LBR364) due to incorrect model labeling, where model 30001 units were labeled as model 30000. The recall affects 30 units distributed nationwide.

    Product
    MEDLINE OPEN HEART, REF DYNJ904261B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0118-2025·2024-10-23

    Beckman Coulter Recalls Immunochemistry IGM Reagent Lots for Sensitivity Failure

    Beckman Coulter is recalling two lots of IGM reagent used in IMMAGE immunochemistry systems because they don't meet sensitivity specifications. The reagent measures immunoglobulin M levels in blood samples.

    Product
    IMMAGE Immunochemistry Systems IGM Immunoglobulin M Reagent REF 447610 IGM reagent, when used in conjunction with IMMAGE Immunochemistry Systems and Calibrator 1, is intended for the quantitative determination of immunoglobulin M (IGM) in human serum by rate nephelometry.
    Category
    Medical Device
    Distribution
    15 states
  • HighFDA (Devices)·Z-0106-2025·2024-10-23

    Boston Scientific AVVIGO+ System Grid Overlay Display Software Error

    Boston Scientific AVVIGO+ Multi-Modality Guidance System displays incorrect grid overlay when connected to OptiCross 18 catheters in Live or Record mode due to a software anomaly. The wrong overlay could affect reference depth settings used in vascular imaging procedures.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL EU + ROW; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0060-2025·2024-10-23

    Monarch Inflation Devices Recalled for Sterile Barrier Defects

    Merit Medical Systems is recalling approximately 6,565 Monarch Inflation Devices due to manufacturing defects that compromise the sterile barrier, potentially allowing device contamination.

    Product
    Monarch Inflation Device 30 atm/bar MAP403 MBA Hemostasis Valve Metal Insertion Tool Torque Device REF IN2430 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0097-2025·2024-10-23

    AVVIGO+ Multi-Modality Guidance System software defect causes incorrect grid display

    Boston Scientific recalled the AVVIGO+ Multi-Modality Guidance System because a software defect causes incorrect grid display in Live and Record modes, potentially affecting vessel imaging guidance for physicians using OptiCross 18 catheters.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + INT SYSTEM - ZERO COST US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0072-2025·2024-10-23

    basixTOUCH Inflation Device May Have Sterile Barrier Defect

    Merit Medical Systems is recalling basixTOUCH Inflation Devices due to a potential hole in the sterile barrier. The defect may expose patients to pathogens and pyrogens, increasing infection risk.

    Product
    basixTOUCH Inflation Device, REF: IN8130/JPC, IN8152/C, IN8302/C, IN8352/C, IN8802/C, IN8130/C, IN8130/C
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0069-2025·2024-10-23

    MONARCH Inflation Device Recalled Due to Sterile Barrier Defect Risk

    Merit Medical Systems recalled 12,910 MONARCH Inflation Devices worldwide due to a potential hole in the sterile barrier. The defect could expose the device to pathogens and pyrogens, potentially leading to infection.

    Product
    MONARCH Inflation Device, REF: IN2430/B, IN2530/B, IN2130/B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0111-2025·2024-10-23

    Boston Scientific AVVIGO+ System Grid Overlay Display Software Error

    Boston Scientific is recalling the AVVIGO+ Multi-Modality Guidance System due to a software error that displays an incorrect 9-grid overlay instead of the correct 15-grid overlay in Live or Record mode, affecting measurement accuracy with OptiCross 18 catheters.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO + MOB Z UPGRADE JP; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0131-2025·2024-10-23

    Rolling Walker Backrest May Disengage During Position Changes Creating Fall Risk

    Nova Ortho-Med is recalling 1,585 MONARCH ROLLATOR RED units because the backrest may disengage from its holder when users adjust from a slouched to upright position, creating a fall hazard.

    Product
    Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product Description: ROLLING WALKER MOBILITY AID
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0058-2025·2024-10-23

    Merit Medical Recalls Monarch Inflation Devices Due to Sterile Barrier Breaches

    Merit Medical Systems is recalling 41,411 Monarch Inflation Devices because manufacturing defects caused holes in the sterile barrier.

    Product
    Monarch Inflation Device MAP 112 Access-9 Large Bore Hemostasis Value Metal Insertion Tool Torque DeviceREF IN2230 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0070-2025·2024-10-23

    Custom Inflation Kits Recalled Due to Sterile Barrier Defect and Infection Risk

    Merit Medical Systems is recalling Custom Inflation Kits due to a small hole in the sterile barrier that could expose users to pathogens and pyrogens, increasing infection risk.

    Product
    Custom Inflation Kit, REF: K05-00052M, K05-00053H, K05-00109G, K05-00137K, K05-00296L, K05-00399D, K05-00404C, K05-00410G, K05-00438G, K05-00502F, K05-00564L, K05-00587D, K05-00612Q, K05-00704F, K05-00724F, K05-00793B, K05-00798, K05-00890D, K05-00890D, K05-00944D, K05-00956D…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0090-2025·2024-10-23

    Leica Biosystems Cryostat Model CM3050 S lacks warning label for flammable sprays

    Leica cryostat units lack a warning about flammable freezing sprays that can ignite in the chamber. The company discovered through post-market surveillance that users were applying these flammable sprays without being warned of the fire hazard.

    Product
    Leica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 14903050S03; for freezing and sectioning tissue samples
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0130-2025·2024-10-23

    Nova Ortho-Med Rollator Backrest May Disengage During Position Adjustment

    Nova Ortho-Med is recalling the MONARCH ROLLATOR PURPLE due to a fall hazard: the backrest may disengage when users shift from a slouched to upright seated position.

    Product
    Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR PURPLE Model/Catalog Number: 4329PL Product Description: ROLLING WALKER MOBILITY AID
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0117-2025·2024-10-23

    Graft Tube shaft may detach from funnel head during surgical use

    Spineology is recalling 190 Graft Tubes because the shaft may detach from the funnel head when not used as instructed, potentially compromising spinal surgical procedures.

    Product
    The Graft Tube is an optional instrument which, as designed, can be filled with flowable graft material, then advanced through the Access Portal. A Push Rod is then used to facilitate the manual deployment of the graft material out of the Graft Tube and adjacent to where the…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0065-2025·2024-10-23

    basixTOUCH Inflation Device Recalled for Sterile Barrier Defect

    Merit Medical Systems is recalling the basixTOUCH Inflation Device (REF: IN8140/A) because devices may have a small hole in the sterile barrier, potentially exposing users to pathogens and pyrogens that could lead to infection.

    Product
    basixTOUCH Inflation Device, REF: IN8140/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0088-2025·2024-10-23

    Leica Cryostat Model CM1900 Lacks Warning on Flammable Sprays

    Leica Biosystems is recalling the Cryostat Model CM1900 because its labeling lacks a warning about using flammable freezing sprays in the chamber. The sprays can ignite, creating a serious fire and burn hazard.

    Product
    Leica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0073-2025·2024-10-23

    Merit Medical StabiliT TOUCH Syringe Sterile Barrier Defect Recall

    Merit Medical Systems is recalling 202 StabiliT TOUCH Syringes with potential small holes in the sterile barrier. Affected devices may expose users to pathogens and pyrogens, which could cause infection.

    Product
    StabiliT TOUCH Syringe, REF: IN8VCF/B
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0125-2025·2024-10-23

    Cortera spinal fixation system screws recalled due to manufacturing specification concerns

    XTANT Medical is recalling 330 Cortera spinal fixation system screws marked 'DNI' because they may not have been manufactured to specification and could lack adequate strength if implanted.

    Product
    Cortera Spinal Fixation System Screw, Poly, Solid, 6.5mm x 45mm, REF: 150100-65045
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0068-2025·2024-10-23

    BlueFIRE Inflation Device Recalled for Sterile Barrier Defect Risk

    BlueFIRE Inflation Devices may have a small hole in the sterile barrier, potentially exposing patients to pathogens that could cause infection. Contact your healthcare provider if you received this device.

    Product
    BlueFIRE Inflation Device, REF: IN10140
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0115-2025·2024-10-23

    Boston Scientific Recalls AVVIGO+ Guidance System Due to Software Anomaly

    Boston Scientific is recalling 4 units of the AVVIGO+ Multi-Modality Guidance System because a software error displays the wrong grid overlay during Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO MOB UPGRADE JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0107-2025·2024-10-23

    Boston Scientific AVVIGO+ Guidance System Grid Mark Overlay Defect

    Boston Scientific is recalling the AVVIGO+ Multi-Modality Guidance System due to a software defect that displays an incorrect 9-grid mark overlay instead of the correct 15-grid mark overlay when used with OptiCross 18 Peripheral Imaging Catheters in Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO + MOB SYSTEM EVAL US; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities which…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0112-2025·2024-10-23

    Medical imaging system software defect causes incorrect grid overlay display

    Boston Scientific recalled the AVVIGO+ imaging guidance system due to a software defect that displays a 9-grid overlay instead of the correct 15-grid overlay when connected to OptiCross 18 catheters in Live or Record mode.

    Product
    Boston Scientific AVVIGO+ Multi-Modality Guidance System, AVVIGO II TO AVVIGO INT UPGRADE JAPAN; the system utilizes an interactive display to provide physicians with physiologic, intravascular and/or intracardiac anatomical information. The system can provide various modalities…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0120-2025·2024-10-23

    Medline Heart Overhead Table Pack recalled for incorrect model labeling

    Medline Industries recalls Heart Overhead Table Packs due to incorrect labeling. Some kits were marked as model 30000 when they should be model 30001.

    Product
    MEDLINE HEART OVERHEAD TABLE PACK, REF DYNJ0778508J
    Category
    Medical Device
    Distribution
    Distributed nationwide